Instructions for use – MicronJetTM
3 Simple Application Steps illustrations

Video Gallery
How to perform a MicronJetTM injection
The beauty of a MicronJetTM injection into the air
Question and Answers
-
What should I do in case of leakage?
- Follow the common practice as with a regular needle and syringe
-
What are the main causes of major leakage?
- Injection may have started out wet – avoid pressing the plunger prior to injection
- Syringe not held properly – make sure you hold the syringe flanges between fingers
- and thumb; avoid switching hands
- Improper device orientation – make sure blue line faces you during injection and
- Syringe is not tilted sideways or rotated
- Too high / low injection angle – make sure angle between syringe axis and skin surface is as close as possible to 45˚
- Insufficient pressure applied to the skin – increase slightly
- Skin not stretched – make sure to keep the skin stretched throughout the entire injection
-
What are the causes of small/no bleb without leakage?
- Large air bubble within syringe – make sure to fill syringe properly and remove air bubbles
- The fluid is leaking between the device and the syringe – make sure you firmly connect the device to the syringe
-
What are the causes of high injection pressure?
- Too much pressure is applied to the skin – reduce slightly
- Improper injection angle – too high. Make sure angle between syringe axis and skin surface is as close as possible to 45˚
- Skin type or texture – try different injection site
Instructions for use – download PDFs:
Go to the product page: MicronJet™ microneedle device